Common Mistakes in Measuring the Fetal Lateral Ventricle

Accurate measurement of the atrium of the lateral ventricle requires strict adherence to the correct imaging plane and caliper placement. Even small technical errors can change the measurement by 1–2 mm, potentially leading to misclassification.

Below are the most common mistakes encountered in clinical practice.

1️⃣ Measuring in the Wrong Plane

The atrial width must be measured in the true transventricular axial plane.

Common errors include:

  • Measuring too superiorly (roof of the ventricle)
  • Measuring too inferiorly (thalami level)
  • Oblique slicing of the head

Oblique sections frequently exaggerate ventricular width.

2️⃣ Not Confirming CSP

The cavum septi pellucidi (CSP) should be clearly visualized in the correct plane.

If the CSP is not visible:

  • The plane may be incorrect
  • The gestational age may be too early
  • There may be associated anomalies

Measurement without confirming the proper plane reduces reliability.

3️⃣ Incorrect Caliper Placement

The measurement must be:

  • Inner edge to inner edge
  • Perpendicular to the long axis of the ventricle
  • Taken at the level of the glomus of the choroid plexus

Common mistakes:

  • Including the outer wall
  • Placing the caliper obliquely
  • Measuring diagonal distance

4️⃣ Including the Choroid Plexus

The choroid plexus should not be included in the measurement.

The atrial width is measured across the ventricular cavity only.

High gain settings may make the choroid plexus appear fused with the wall, leading to underestimation or overestimation.

5️⃣ Measuring Only One Side

When the measurement is borderline (9–10 mm):

  • Both ventricles should be assessed
  • Re-measure after slight probe adjustment
  • Confirm symmetry

Small angulation differences can change the result significantly.

6️⃣ Overreacting to a Single Borderline Measurement

A measurement of 10 mm does not automatically indicate pathology.

Proper follow-up and repeat measurement are often more important than immediate labeling.

Clinical Tip

Before reporting mild ventriculomegaly, always:

✔ Confirm the correct plane

✔ Re-measure after adjusting probe angle

✔ Ensure inner-to-inner caliper placement

✔ Check both sides

Precision reduces unnecessary parental anxiety.

글쓴이

UltraLog

I share practical fetal ultrasound knowledge based on real clinical experience.